A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure

Patel, Darshan Shyam

Abstract Details

2008, Master of Science in Engineering, University of Akron, Biomedical Engineering.
In critical care, a fluid-filled catheter transducer system is commonly used to continuously monitor blood pressure. The fluid-filled catheter transducer system, as a first approximation, behaves as a second order dynamic system. The dynamic characteristics of the system are affected by variations in assembly technique and time dependent changes in the system, often resulting in distortion and inaccurate measurements. A previous simulation study employed a counter pressure source in tandem with the transducer. The counter pressure generated minimized the fluid flow in the pressure monitoring system. When the flow in the system was zero the counter pressure generated closely approximated the true blood pressure. The current study developed a real-time technique to generate accurate and dependable counter pressure. To validate this technique, one experimental model (second order dynamic system) and two simulation models (second order dynamic system and fourth order dynamic system) of catheter transducer system were developed and tested under varying system conditions. The real-time technique developed, successfully reproduced the true blood pressure waveforms, regardless of variations in the system characteristics and changes in the system over time.
Bruce Taylor, PhD (Advisor)
Daniel Sheffer, PhD (Committee Member)
Dale Mugler, PhD (Committee Member)

Recommended Citations

Citations

  • Patel, D. S. (2008). A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure [Master's thesis, University of Akron]. OhioLINK Electronic Theses and Dissertations Center. http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260

    APA Style (7th edition)

  • Patel, Darshan. A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure. 2008. University of Akron, Master's thesis. OhioLINK Electronic Theses and Dissertations Center, http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260.

    MLA Style (8th edition)

  • Patel, Darshan. "A Real-Time Technique for the Correction of Invasive Blood Pressure Measurements using Counter Pressure." Master's thesis, University of Akron, 2008. http://rave.ohiolink.edu/etdc/view?acc_num=akron1205764260

    Chicago Manual of Style (17th edition)