Skip to Main Content
 

Global Search Box

 
 
 
 

ETD Abstract Container

Abstract Header

MX908®: Sensitivity and Limit of Detection Evaluation of On Swab and Off Table Samples

Abstract Details

2022, Master of Science (MS), Bowling Green State University, Forensic Science.
908 Devices has created a handheld High-Pressure Mass Spectrometer™ (HP-MS), called the MX908®, for field analysis of unknown substances. According to the manufacturer, the MX908® has a theoretical limit of detection in the low to mid parts per billion (ppb) range, but because the device provides qualitative rather than quantitative data, the actual limits of detection are unknown. The purpose of this experiment was to determine a limit of detection and sensitivity of the MX908® for methamphetamine, fentanyl, heroin, and cocaine by comparing known standards for each drug processed on the MX908® and a Shimadzu 8050 triple quadrupole Liquid Chromatograph Mass Spectrometer (LC-MS). Samples were spotted directly on the MX908® swabs and spotted on a table and swabbed off. This was to compare a controlled setting to a simulation of a practical use of the swabs and instrument. Cocaine and methamphetamine were the only two drugs reliably detected by the instrument. More testing is necessary to validate the MX908® for future use
Travis Worst, Ph.D (Committee Chair)
Jeremy Canfield, MS (Committee Member)
Jon Sprague, Ph.D (Committee Member)
29 p.

Recommended Citations

Citations

  • Brown, W. (2022). MX908®: Sensitivity and Limit of Detection Evaluation of On Swab and Off Table Samples [Master's thesis, Bowling Green State University]. OhioLINK Electronic Theses and Dissertations Center. http://rave.ohiolink.edu/etdc/view?acc_num=bgsu1661102197749377

    APA Style (7th edition)

  • Brown, Wyatt. MX908®: Sensitivity and Limit of Detection Evaluation of On Swab and Off Table Samples. 2022. Bowling Green State University, Master's thesis. OhioLINK Electronic Theses and Dissertations Center, http://rave.ohiolink.edu/etdc/view?acc_num=bgsu1661102197749377.

    MLA Style (8th edition)

  • Brown, Wyatt. "MX908®: Sensitivity and Limit of Detection Evaluation of On Swab and Off Table Samples." Master's thesis, Bowling Green State University, 2022. http://rave.ohiolink.edu/etdc/view?acc_num=bgsu1661102197749377

    Chicago Manual of Style (17th edition)