Validation of LeadCare (c) II Portable Blood Lead Analyzer
Author Info
Kuyat, Shelly
Permalink:
http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171
Abstract Details
Year and Degree
2014, MS, University of Cincinnati, Medicine: Industrial Hygiene (Environmental Health).
Abstract
This study compares a LeadCare © II blood lead analyzer with a LeadCare © I blood lead analyzer and determines the correlation of the analytical results. Subjects recruited for a blood lead level monitoring study during a building demolition/renovation project participated in this study. A total of 12 eligible study participants were recruited. Two background samples (background 1 and background 2) were collected via venipuncture. Background 1 sample results from the LeadCare © I analyzer ranged from < 1.4 to 4.1 µg/dL. The background 1 sample mean was 1.75 µg/dL with a SD ± 0.80. Results for the background 2 sample ranged from < 1.4 to 2.7 with a mean of 1.53 µg/dL and a SD ± 0.39. The lower reportable range of the LeadCare © II is 3.3 µg/dL and 1.4 µg/dL for the LeadCare © I. Most of the sample results reported by the LeadCare © I were less than 3.3 µg/dL making a comparison with the LeadCare © II impractical.
Committee
Glenn Talaska, Ph.D. (Committee Chair)
Nicholas Newman, D.O. M.S. (Committee Member)
Michael Maier, Ph.D.M.S. (Committee Member)
Pages
32 p.
Subject Headings
Occupational Safety
Keywords
LeadCare
;
Portable blood lead analyzer
Recommended Citations
Refworks
EndNote
RIS
Mendeley
Citations
Kuyat, S. (2014).
Validation of LeadCare (c) II Portable Blood Lead Analyzer
[Master's thesis, University of Cincinnati]. OhioLINK Electronic Theses and Dissertations Center. http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171
APA Style (7th edition)
Kuyat, Shelly.
Validation of LeadCare (c) II Portable Blood Lead Analyzer.
2014. University of Cincinnati, Master's thesis.
OhioLINK Electronic Theses and Dissertations Center
, http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171.
MLA Style (8th edition)
Kuyat, Shelly. "Validation of LeadCare (c) II Portable Blood Lead Analyzer." Master's thesis, University of Cincinnati, 2014. http://rave.ohiolink.edu/etdc/view?acc_num=ucin1407410171
Chicago Manual of Style (17th edition)
Abstract Footer
Document number:
ucin1407410171
Download Count:
788
Copyright Info
© 2014, all rights reserved.